September 30, 1994
Design Standards
EtO Sterilizer/Aerator Room
ARCHITECTURAL
Ceiling
PCP
Wainscot
-
Ceiling Height
2700 (9'-0")
Base
-
Wall Finish
PRE-FAB PANEL
Floor Finish
C
Slab Depression
-
SPECIAL EQUIPMENT
VS-325: ETO STERILIZER
VS-360: ETO DISPOSER SYSTEM
ELECTRICAL
Lighting
Power
General
50 FC - 3 W/SF*
General
***
Special
30 FC-2 W/SF**
Emergency
***
Notes: *LOBBY
**SERVICE AREA
***VS-325: POWER DOOR - 17A, 120V; CONTROL - 3A, 120V;
VS-360: 40A, 208V, 3 PH
***VACUUM PUMP - 8A, 208V, 3 PH;
ADP
-
Telephone
-
Radio
-
Intercom
-
Public Address -
HEATING, VENTILATING AND AIR CONDITIONING
Dry Bulb Temp Cooling F
76
Relative Humidity/Cooling -
Minimum % Outside Air
100
Minimum Air Changes/Hr
10
YES
Room Pressure
NEGATIVE
Noise Criteria
NC-40
AC Load - Equipment
2.0438 W/SM
Steam
-
Number of People
0
Notes:
Special Exhaust
YES****
****DEDICATED ETO EXHAUST SYSTEM PER HVAC DESIGN MANUAL
VS-360 WHERE REQ'D TO MEET STATE CODES
PLUMBING AND MEDICAL GASES
Cold Water
YES
Sanitary Drain
YES*****
Hot Water
-
Reagent Grade Water
-
Laboratory Air
-
Medical Air
-
Laboratory Vacuum
-
Medical Vacuum
-
Other
SHOP AIR******
Oxygen
-
Notes: *****4" OSD FOR EACH STERILIZER
******FOR VS EQUIP
VA Design Guide -- Supply, Processing, & Distribution
Guide Plate 5-3