September 30, 1994
Design Standards
Manual Equipment Wash Area
ARCHITECTURAL
Ceiling
PCP
Wainscot
-
Ceiling Height
2700 (9'-0")
Base
QT
Wall Finish
CT*
Floor Finish
QT
Notes:
Slab Depression
-
*MOUNT CT ON CONCRETE BACKER BOARD
ELECTRICAL
Lighting
Power
General
50 FC - 3 W/SF
General
Special
-
Emergency
-
ADP
-
Telephone
-
Radio
-
Intercom
-
Notes:
Public Address -
HEATING, VENTILATING AND AIR CONDITIONING
Dry Bulb Temp Cooling F
-
Relative Humidity/Cooling -
Minimum % Outside Air
-
Minimum Air Changes/Hr
10
YES
Room Pressure
NEGATIVE
Noise Criteria
NC-40
AC Load - Equipment
-
Steam
YES
Number of People
0
Notes:
Special Exhaust
YES**
**DEDICATED EXHAUST SYSTEM, SS CONSTRUCTION, FOR SATURATED WET EXHAUST AIR
PLUMBING AND MEDICAL GASES
Cold Water
YES
Sanitary Drain
YES***
Hot Water
YES
Reagent Grade Water
-
Laboratory Air
-
Medical Air
-
Laboratory Vacuum
-
Medical Vacuum
-
Other
SHOP AIR****
Oxygen
-
Notes:
***DRAIN AT ROOM CENTER; SLOPE FLOOR AT 1/8" PER FT
****TO ROOM
VA Design Guide -- Supply, Processing, & Distribution
Guide Plate 5-4