Final Draft 12-31-95
Revised May 1997
Equipment Guide List
Special Procedure Room (non-affiliated Control Room)
SYMBOL
QTY AI
DESCRIPTION
NOTE: THE FOLLOWING X-RAY EQUIPMENT TO BE DETERMINED ON PROJECT BASIS BY
VETERANS HEALTH ADMINISTRATION.
NOTE: SHOULD VETERANS HEALTH ADMINISTRATION INDICATE A SINGLE PLANE
ANGIOGRAPHIC FILM CHANGER, THE RADIOGRAPHIC UNIT FOR THE SECOND PLANE
WILL BE OMITTED.
NOTE: OPTIONAL COMPONENTS TO BE DETERMINED BY VETERANS HEALTH
ADMINISTRATION.
NOTE: OUTLET, WALL, EVACUATION (PG-18-1, MCS 15491; H-18-3, CS 310-1))
NOTE: OUTLET, WALL, MEDICAL AIR (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: OUTLET, WALL, NITROUS OXIDE (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: OUTLET, WALL, OXYGEN (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: OUTLET, WALL, VACUUM (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: OUTLETS TO BE ON THE WALL, NEAR HEAD END AND OPPOSITE END OF X-RAY
TABLE, 1524mm (60") ABOVE THE FLOOR.
NOTE: SERVICES, ELECTRICAL, AS REQUIRED FOR EQUIPMENT.
NOTE: ISOLATED POWER UNIT PROVIDES ISOLATED ELECTRICAL POWER, INCLUDES
LINE ISOLATION MONITOR, ISOLATION TRANSFORMER AND CIRCUIT BREAKERS (DETAILS
AVAILABLE FROM ELECTRICAL SERVICE)
NOTE: LOW LEVEL DIMMED INCANDESCENT LIGHTING SYSTEM IN ADDITION TO GENERAL
NOTE: SHIELDING, RADIATION, FOR ROOMS WITH
FIXED X-RAY EQUIPMENT, IN
ACCORDANCE WITH (PG-18-1, MCS 13091; H-18-3, CS 64-1; PG-18-4, SD 24 THRU 24K AND
NCRP REPORT NO. 33, 35 AND 49)
AR
CF
MODULAR CASEWORK
J-2
1
CF
SINK, CORROSION RESISTING STEEL, WITH CENTER DRAIN
OUTLET, 457mm x 381mm 203mm (18" x 15" x 8")
(PG-18-1 & PG-18-6, MCS 11602)
1
VC
X-RAY UNIT, RADIOGRAPHIC AND FLUOROSCOPIC
CONSISTS OF THE FOLLOWING COMPONENTS:
1
VC
CAMERA, 35/100 OR 105mm, USE WITH IMAGE INTENSIFIER
(OPTIONAL)
1
VC
CAMERA, TV (FOR USE WITH IMAGE INTENSIFIER) (OPTIONAL)
1
VC
CARRIAGE, TUBE, CEILING MOUNTED, WITH TWO TUBE CRANES
1
VC
CHANGER, FILM, ANGIOGRAPHIC, BI-PLANE
1
VC
CHANGER, FILM, DUAL RAPID
1
VC
CONTROL, CINE CAMERA
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 4-4