Final Draft 12-31-95
Revised May 1997
Design Standards
CT Suite - Mid-Level Imaging System
ARCHITECTURAL
POWER (cont.)
Ceiling:
Notes: 2. Electrical trades shall provide
Ceiling Ht:
2890 mm (9'-6")
necessary conduit, openings,
Wall Finish:
--
bushings, nipples, flexible conduit
Wainscot:
--
etc., as required at the various
Base:
Resilient base
junction boxes, duct and conduit
Floor Finish:
terminations to allow proper
Access Floor System in
connections of the x-ray equipment
Computer and Control Rooms
and related accessories.
Slab Depr:
100mm (4")
3. 480V., 3P., 150A., adjustable trip
Notes: 1. 1200 mm (4'-0") wide door. See VA
flush mounted circuit breaker with
Construction Standard 64-1, X-Ray
shunt trip. Run empty 50 mm (2') C.
from circuit breaker to x-ray duct
Control".
above finished ceiling.
4. 300 mm x 140 mm (12' x 5-") deep
SPECIAL EQUIPMENT
flush floor duct with screw-on cover.
None
Connect to x-ray wall duct.
5. Run 13 mm (") C. from door switch
LIGHTING
to junction box at x-ray controller.
Connect door switch so that if door is
General:
30 FC (CT Gantry Room)
opened the x-ray system will be shut
30 FC (Control Room)
off.
Special:
30 FC (Computer Room)
Notes: 1. 600 mm x 1200 mm (2'x4') recessed
COMMUNICATION
fluorescent light fixture, acrylic
prismatic lens, W/3-F32T8 lamps,
ADP:
Yes
3500 deg K, CRI-75 (minimum)
Radio:
--
Telephone:
Yes
downlight with 100W, A-21 lamp.
Intercom:
Yes
Pub. Addr:
--
Notes: 1. Junction box for CCTV camera with
controlled by dimmer.
conduit to adjacent control room.
2. Junction box for CCTV monitor.
POWER
General:
Emergency:
--
Notes: 1. The electrical plan as shown is to be
used as a guide only. Equipment
locations, dimensions and wiring
requirements to be per the x-ray
system suppliers equipment
installation drawings.
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 4-5