Final Draft 12-31-95
Revised May 1997
Design Standards
Special Procedures (non-affiliated Control Room)
ARCHITECTURAL
POWER (cont.)
Ceiling:
Notes: 2. Electrical trades shall provide
Ceiling Ht:
Coordinate/ with equipment
necessary conduit, openings,
Manufacturer
bushings, nipples, flexible conduit
Wall Finish:
Painted veneer plaster
etc., as required at the various
Wainscot:
--
junction boxes, duct and conduit
Base:
Integral welded seam sheet
terminations to allow proper
flooring base
connections of the x-ray equipment
Floor Finish:
Welded seam sheet flooring
and related accessories.
Slab Depr:
100 mm (4")
3. 480V., 3P., 150A., adjustable trip
Notes: 1. 1200 mm (4'-0") wide door. See VA
flush mounted circuit breaker with
Construction Standard 64-1, X-Ray
shunt trip. Run empty 50mm (2") C.
from circuit breaker to x-ray duct
Control".
above finished ceiling.
4. 300 mm x 140 mm (12' x 5-") deep
SPECIAL EQUIPMENT
flush floor duct with screw-on cover.
None
Connect to x-ray wall duct.
5. 250 mm x 140 mm (10" x 5-") deep
LIGHTING
flush vertical wall duct with 305mm
(12") wide screw-on cover. Connect
General:
50 FC (Procedure)
to floor duct and to x-ray duct above
Special:
10 FC (Control)
finished ceiling.
Notes: 1. X-ray in use DO NOT ENTER sign
6. 250 mm x 140 mm (10" x 5-") deep
fixture
flush vertical wall duct with 300 mm
2. Continuous fluorescent direct/indirect
(12") wide screw-on cover. Connect
wall mounted light fixture, acrylic
to x-ray duct above finished ceiling
prismatic lens on top and bottom,
and terminate at finished floor.
W/2-F32T8 lamps, 3500 deg. K,
7. 250 mm x 140 mm (10" x 5-") deep
CRI=75 (minimum.)
x-ray duct above finished ceiling with
255mm (10") wide screw-on cover.
downlight with 100W, A-21 lamp.
Connect to vertical wall duct.
8. Emergency Power Off (EPO) push
fluorescent light fixture, acrylic
button station. Connect to shunt trip
prismatic lens, W/3-F32T8 lamps,
circuit breakers.
3500 deg. K, CRI-75 (minimum).
controlled by dimmers.
POWER
General:
Emergency:
--
Notes: 1. The electrical plan as shown is to be
used as a guide only. Equipment
locations, dimensions and wiring
requirements to be per the x-ray
system suppliers equipment
installation drawings.
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 4-4