Final Draft 12-31-95
Revised May 1997
Equipment Guide List
Radiographic Fluoroscopic Room w/Control
SYMBOL
QTY AI
DESCRIPTION
NOTE: PROVIDE PIPING CONNECTION FROM PROCESSORS TO SILVER RECOVERY
SYSTEM, IF REQUIRED.
NOTE: SERVICES, ELECTRICAL, AS REQUIRED FOR EQUIPMENT.
NOTE: PROVIDE ONLY WHEN APPROVED FOR A SPECIFIC PROJECT (OXYGEN, AIR,
VACUUM OUTLET AND BRACKET)
.
NOTE: OUTLETS TO BE ON THE WALL, NEAR HEAD END OF X-RAY TABLE, 1524mm (60")
ABOVE THE FLOOR.
NOTE: OUTLET, WALL, MEDICAL AIR (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: OUTLET, WALL, OXYGEN (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: OUTLET, WALL, VACUUM (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: CONNECTIONS, PLUMBING, ELECTRICAL OR MECHANICAL AS REQUIRED
NOTE: LOW LEVEL DIMMED INCANDESCENT LIGHTING SYSTEM IN ADDITION TO GENERAL
NOTE: SHADES, EXTERIOR WINDOWS (PG-18-1, MCS 12513)
NOTE: SHIELDING, RADIATION, FOR ROOMS WITH
FIXED X-RAY EQUIPMENT, IN
ACCORDANCE WITH (PG-18-1, MCS 13091; H-18-3, CS 64-1; PG-18-4, SD 24 THRU 24K AND
NCRP REPORT NO. 33, 35 AND 49)
1
CC
DISPENSER, SOAP, WALL MOUNTED WITH FOOT CONTROL
(PG-18-1, MCS 10800)
P-418
1
CC
LAVATORY, SENSOR CONTROL (PG-18-1, MCS 15450, VOL. 3
SD 380)
1
CF
CABINET, STORAGE, WALL MOUNTED, 254mm (10") ABOVE
FINISHED FLOOR, 5 SHELVES AND FOLDING DOOR,
APPROX., 610mm x 406mm x 1575mm (24" x 16" x 62")
1
VC
X-RAY UNIT, RADIOGRAPHIC AND FLUOROSCOPIC CONSISTS OF
THE FOLLOWING COMPONENTS:
1
VC
BUCKY, TILT, WALL MOUNTED (USE IN TYPE R-F #2 AND TYPE
R-F #3)
1
VC
CAMERA UNIT, CINE, FOR USE WITH IMAGE INTENSIFIER
(OPTIONAL)
1
VC
CAMERA, SPOT, 100mm OR 105mm (USE IN TYPE R-F #3)
1
VC
CAMERA, TV (FOR USE WITH IMAGE INTENSIFIER) (OPTIONAL)
1
VC
CHANGER, CASSETTE
1
VC
CONTROL, CINE CAMERA
1
VC
IMAGE INTENSIFIERM 152mm, 229mm (6", 9"), OR DUAL TYPE TO
BE DETERMINED ON PROJECT BASIS
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 4-3