Final Draft 12-31-95
Revised May 1997
Design Standards
Radiographic Fluoroscopic Room w/Control
ARCHITECTURAL
POWER (cont.)
Notes:
3. 480V., 3P., 150A., adjustable trip
Ceiling:
flush mounted circuit breaker with
Ceiling Ht:
2890 mm (9'-6")
shunt trip. Run empty 600 mm (2')
Wall Finish:
Paint
C. from circuit breaker to x-ray
Wainscot:
--
duct above finished ceiling.
Base:
Resilient base
4. 240V, 2P., 30A flush MTD. circuit
Floor Finish:
breaker for film processor.
Slab Depr:
100 mm (4")
5. 250 mm x 140 mm (10" x 5-")
Notes:
1200 mm (4'-0") wide door. See VA
deep flush vertical wall duct with
Construction Standard 64-1, X-Ray
300mm (12") wide screw-on
cover. Connect to floor duct and
Control".
to x-ray duct above finished
ceiling.
SPECIAL EQUIPMENT
6. 250 mm x 140 mm (10" x 5-")
None
deep flush vertical wall duct with
300 mm (12") wide screw-on
LIGHTING
cover. Connect to x-ray duct
General:
30 FC - 2.0W/sf
above finished ceiling and
Special:
10FC - 3.0W/sf-
terminate at finished floor.
Notes:
1. Continuous fluorescent
7. 250 mm x 140 mm (10" x 5-")
direct/indirect wall mounted fixture,
deep x-ray duct above finished
acrylic prismatic lens on top and
ceiling with 250 mm (10") wide
bottom, W/1-F32T8 lamp 3500
screw-on cover. Connect to
deg. K, CRI=75 (min.)
vertical wall duct.
2. "X-ray in use DO NOT ENTER"
8. Emergency Power Off (EPO)
sign fixture.
push button station. Connect to
shunt trip breakers.
downlight with 100W, A-21 lamp.
4. Provide dimmers for down light
COMMUNICATION
ADP:
--
POWER
Radio:
--
Telephone:
Yes
General:
Intercom:
Yes
Emergency:
--
Pub. Addr
--
Notes: 1.The electrical plan as shown is to be
Notes:
1. Nurse Call in toilet room to
used as a guide only. Equipment
annunciate at reception desk.
locations, dimensions and wiring
requirements to be per the x-ray
system suppliers equipment
installation drawings.
2.Electrical trades shall provide
necessary conduit, openings,
bushings, nipples, flexible conduit
etc., as required at the various
junction boxes, duct and conduit
terminations to allow proper
connections of the x-ray equipment
and related accessories.
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 4-3