HEATING, VENTILATING AND AIR CONDITIONING
ARCHITECTURAL
AC Load Lights
8.5 W/SF
Floor Area
450 NSF (41.85 NSM) Wall Finish GYP. BOARD (SC)
ACROVYN ON CBB
Ceiling
GYP. BOARD
Wainscot
AC Load Equipment
16.0 W/SF
10'-0"* (3.0 METERS) Base
6" (152 mm) INTEGRAL
Number of People
12
Ceiling Height
Noise Criteria
NC-40
Floor Load
100 PSF
COVE BASE
Floor finish WSF
Room Pressure
POSITIVE
Note:
Lead Lining AS REQUIRED
62 - 80F (17 - 27C)
Dry Bulb Temp Cooling Range
Refer to PG-18-1 and PG-18-6
62 - 80F (17 - 27C)
*ADD 8" ACCESSIBLE SPACE ABV CLG FOR MICROSCOPE
Minimum Air Changes per Hour
15 OCC/8 UNOCC
OR 10'-2" (3.05 METERS)
Minimum% Outside Air
100
%
YES
ELECTRICAL
-
Steam
-
Lighting
Power
Relative Humidity
45-55 %
Relative Humidity
45-55 %
General
200 FC, 6.0 W/SF*
General
(1) MODULE EA WALL
Special
SURGICAL LIGHT**
(1) MODULE EA COLUMN
Emergency
(4) FIXTURES***
Notes:
Special
****
Emergency*****
PLUMBING AND MEDICAL GASES
*COLOR IMPROVED FLUOR LAMPS MATCHING COLOR TEMPERATURE OF
SURGICAL LIGHT
YES
YES
Cold Water
Medical Air
**(1) TYPE A, 1000 W
FLUSHING
YES
Hot Water
Medical Vacuum
RIM FLR.
***BATTERY BACKUP IN (4) FLUOR FIXTURES
Oxygen
YES
DRAIN
-
-
****(2) 5 KVA 10-CIRCUIT IPU
Sanitary Drain
YES
Nitrous Oxide
YES
*****EACH IPU & X-RAY UNIT, (1) FILM PROCESSOR PER SUITE
-
Acid Waste
YES
Silver Recovery
-
Anesthesia Evac
YES
TELECOMMUNICATIONS
Data
WALL TERMINAL @
Patient Monitor
YES
EACH UTILITY COLUMN
-
Nurse Call
SPECIAL EQUIPMENT
Telephone
WALL MTD HAND FREE @
Code One
-
EACH UTILITY COLUMN
EMPTY
CCTV
None
CONDUIT
COMB. W/TELEPHONE
Intercom
Public Addr.
EMPTY CONDUIT
ADP
EMPTY CONDUIT
Radio
EMPTY CONDUIT
Design Guide - Surgical Service
Design Standards
Cystoscopy Room
Guide Plate:
6-1h
450 NSF (41.85 NSM)
Date: August 2005