ARCHITECTURAL
HEATING, VENTILATING AND AIR CONDITIONING
Floor Area
120 NSF (11.0 NSM)
Wall Finish
GYP. BOARD(SC)
AC Load Lights
1.5 W/SF
-
Ceiling
GYP. BOARD
Wainscot
AC Load Equipment
7.5 W/SF
Base
9'-0" (2.75 METERS)
WSF
Ceiling Height
Number of People
1
Floor Load
100 PSF
6" (152 mm) INTEGRAL
Floor finish
Noise Criteria
NC-40
COVE BASE
Room Pressure
NEGATIVE
Lead Lining
-
Dry Bulb Temp Cooling Range
78 F (25 C)
Note:
72 F (22 C)
Refer to PG-18-1 and PG-18-6
Minimum Air Changes per Hour
8(SA)
Minimum % Outside Air
100
YES
4" ROUND DUCT*
ELECTRICAL
Steam
-
Relative Humidity
50 %
Lighting
Power
Relative Humidity
30 %
General
40 FC, 1.5 W/SF*
General
360 WATTS
Special
0-10 FC, 0.5 W/SF*
Special
-
*FOR FILM PROCESSOR
SAFETY LIGHT*
Emergency
-
DUCTS REQUIRE LEAD SHIELDING
-
Emergency
*INCANDESCENT
PLUMBING AND MEDICAL GASES
YES
-
Cold Water
Medical Air
YES
-
Hot Water
Medical Vacuum
Sanitary Drain
-
Oxygen
-
YES
Acid Waste
Nitrous Oxide
-
Silver Recovery
YES
TELECOMMUNICATIONS
-
Anesthesia Evac
-
Patient Monitor
-
Telephone
-
-
-
Nurse Call
Intercom
Code One
-
Public Addr.
-
-
CCTV
ADP
-
SPECIAL EQUIPMENT
Radio
-
None
Design Guide - Surgical Service
Design Standards
Radiographic Film
Guide Plate:
5-5b
Processing Room
120 NSF (11.0 NSM)
Date: August 2005