Spinal Cord Injury Center Design Guide
July 1990 (Rev. Jan. 2004)
Cystoscopic Room/Control Alcove 28-A
Equipment Table:
SYMBOL
QUANTITY
AI
DESCRIPTION
A
B
C
D
1
VC
TABLE, UROLOGICAL RADIOGRAPHIC, MOTOR DRIVEN, WITH X-RAY TUBE
SUPPORT
1
CC
DRAIN, FLOOR, AUTOMATIC FLUSHING TYPE (PG-18-1, MCS 15400; PG-18-4,
SD 15400-1.DWG)
1
VC
RADIOGRAPHIC TUBE AND HIGH VOLTAGE CABLES, 500 MA, SINGLE PHASE
1
VC
TRANSFORMER, RADIOGRAPHIC, FOR USE WITH UROLOGICAL TABLE, 50 KVA
1
VV
CLOCK, ELECTRIC TIME ELAPSED
NOTE: THE FOLLOWING EQUIPMENT IS PROVIDED FOR THE CONTROL ROOM
(WHEN TWO CYSTOSCOPIC ROOMS ARE PROVIDED FOR THE PROJECT A
COMMON CONTROL ROOM WILL BE ACCEPTABLE, AND A SINGLE X-RAY
GENERATOR WILL BE INSTALLED TO ENERGIZE BOTH CYSTOSCOPIC TABLES.
AN ADDITIONAL LEAD GLASS VIEWING WINDOW WILL BE REQUIRED FOR THE
SECOND ROOM. WHERE TWO CYSTOSCOPIC TABLES ARE INSTALLED IN ONE
BETWEEN TABLES.)
1
CC
WINDOW, VIEWING, LEAD GLASS, FOR PATIENT OBSERVATION (PG-18-1,
MCS 13091)
1
VC
CONTROL CABINET
1
CC
STATION, INTERCOM, IN CONTROL ROOM SPEAKER IN CYSTOSCOPIC ROOM
(PG-18-1, MCS 16760)
1
CC
OUTLET, INTERCOM (EMPTY CONDUIT SYSTEM) (PG-18-1, MCS 16111)
A
CC
SERVICES, ELECTRICAL, SPECIAL AS REQUIRED FOR THE ABOVE EQUIPMENT
R
(THE X-RAY TRANSFORMER MAY BE LOCATED OUTSIDE THE CYSTOSCOPIC
ROOM, PROVIDED THE DISTANCE DOES NOT EXCEED THAT PRESCRIBED BY
THE EQUIPMENT MANUFACTURERS. X-RAY TRANSFORMERS SHOULD NOT BE
SUSPENDED FROM CEILING OR WALL MOUNTS)
A
CC
SHIELDING, RADIATION, FOR ROOMS WITH FIXED X-RAY EQUIPMENT, IN
R
ACCORDANCE WITH (PG-18-1, MCS 13091; H-08-3, CS 64-1; H-08-4, SD 24, 24K
AND NCRP REPORT NO. 49)
A
CC
RECEPTACLE, ELECTRICAL, DUPLEX, 120 VOLT, 20 AMP (PG-18-1, MCS 16140)
R
1
VV
RECEPTACLE, WASTE, STEP ON TYPE, APPROX. 12" DIAMETER
1
CC
ISOLATED POWER UNIT PROVIDES ISOLATED ELECTRICAL POWER, INCLUDES
LINE ISOLATION MONITOR, ISOLATION TRANSFORMER AND CIRCUIT
BREAKERS (DETAILS AVAILABLE FROM ELECTRICAL SERVICE)
TABLE IV-42
CYSTOSCOPIC ROOM/CONTROL ALCOVE
Office of Facilities Management
156