FINAL DRAFT 2-29-96
Revised 10/96
Equipment Guide List
Mold Room
SYMBOL
QTY AI
DESCRIPTION
NOTE: CEILING LIGHTING SHOULD BE LOCATED TO SUIT BENCH LOCATIONS AND
ORIENTATIONS.
NOTE: CONNECTIONS, PLUMBING, ELECTRICAL OR MECHANICAL AS REQUIRED
NOTE: RECEPTACLE, ELECTRICAL, PROVIDE BOTH 120 & 208 VOLT SERVICE, STRIP,
MULTIPLE, HEAVY DUTY, 1067 mm (42") ABOVE FINISHED FLOOR ON
WALLS W/OUTLET AT 914 mm (36") INTERVALS (H-18-1, MCS 16140; H-18-3, CS 801-3)
AR
CC
HOOD, CANOPY, EXHAUST, PROVIDED OVER POWER
TOOLS AND MELTING POWER AREAS, DIMENSIONS TO
BE INDICATED ON DRAWINGS (H-18-1, MCS 15822)
1
CC
SINK, CORROSION RESISTING STEEL, DOUBLE
COMPARTMENT, ELBOW CONTROLS
AR
CF
MODULAR CASEWORK
1
VV
DRILL PRESS, BENCH MOUNTED
1
VV
DUST COLLECTOR, MOBILE
1
VV
GRINDER, BENCH MOUNTED
1
VV
ILLUMINATOR, X-RAY FILM, 2 PANEL, SURFACE MTD.
1
VV
MILL, COMPENSATOR
1
VV
PROJECTOR, COMPENSATOR
1
VV
PUMP, SUCTION, MOBILE
1
VV
SAW, BAND, BENCH TYPE
AR
VV
STOOL, ADJUSTABLE HEIGHT, WITHOUT ARMS
1
VV
STYROFOAM CUTTER
1
VV
TABLE, BLOCK COOLING
AR
VV
WORK BENCH, WOOD TOP WITH STORAGE
UNDERNEATH, APPROX., 1219 mm (48") X LENGTH AS
REQUIRED X 762 mm (30")
1
VV
DISPENSER, PAPERTPWEL, SURFACE MOUNTED
1
VV
DISPENSER, SOAP, LIQUID, WALL MOUNTED
VA DESIGN GUIDE
GUIDE PLATE 5-1