Final Draft 12-31-95
Revised May 1997
Design Standards
ARCHITECTURAL
CONDITIONING
Ceiling:
Ceiling Ht:
2700 mm (9'-0")
Dry Bulb Temp Cooling:
25C (78F)
Wall Finish:
Paint on Gypsum Wallboard
22C (72F)
Wainscot:
--
Minimum % Outside Air:
100%
Base:
Resilient base
Yes
Floor Finish:
Noise Criteria:
NC-50
Slab Depr:
--
Steam:
--
Notes:
--
Relative Humidity/Cooling:
50%
Relative Humidity/Heating:
30%
SPECIAL EQUIPMENT
Minimum Air Changes/Hr.:
Room Pressure:
Negative
None
AC Load-Equipment:
65 W/sm (6W/sf)
AC Load Lighting:
16 W/m 2 (1.5 W /sf)
LIGHTING
Number of People:
6
General:
30 FC - 1.5w/sf
Special Exhaust:
--
Special:
--
Notes: 1. Provide an exhaust duct connection
Notes: 1. 600 mm x 1200 mm (2' x 4')
to each processor.
acrylic prismatic lens, W/2-F32T8
PLUMBING AND MEDICAL GASES
lamps, 3500 deg. K, CRI=75
Cold Water:
Yes
(minimum).
Hot Water:
Yes
Laboratory Air:
--
fixture.
Laboratory Vacuum:
--
Sanitary Drain:
Yes
POWER
Reagent Grade Water:
--
General:
Medical Air:
--
Emergency:
--
Medical Vacuum:
--
Oxygen:
--
Notes: 1. 240V, 2P., 30A, flush mounted circuit
Notes: 1. Check with film processor supplier
breaker for film processor (coordinate
regarding water quality supplied to
requirements with system supplier).
each processor.
2. Provide piping connection from
COMMUNICATION
processors to silver recovery system.
ADP:
--
if required.
Radio:
--
Telephone:
Yes
Intercom:
--
Pub. Addr:
--
Notes:
--
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 5-4