Final Draft 12-31-95
Revised May 1997
Design Standards
Patient Gowning (for 4 Diagnostic Rooms)
ARCHITECTURAL
CONDITIONING
Ceiling:
Ceiling Ht:
2700 mm (9'-0")
Dry Bulb Temp Cooling:
25C (78F)
Wall Finish:
Paint
22C (72F)
Wainscot:
--
Minimum % Outside Air:
15%
Base:
Resilient base
For Toilet Room
Ceramic Tile in Toilet Room
Noise Criteria:
NC-40
Floor Finish:
Steam:
--
Ceramic Tile in Toilet Room
Relative Humidity/Cooling:
50%
Slab Depr:
--
Relative Humidity/Heating:
30%
Notes:
--
Minimum Air Changes/Hr. Toilet Room exhaust 10
All Others: 4
SPECIAL EQUIPMENT
Room Pressure:
Toilet Room: Negative
All Others: 0
None
AC Load-Equipment:
26 W/m2 (2.4W/sf)
AC Load Lighting:
LIGHTING
Toilet Room: 15W/m2 (1.4 W/sf)
General:
30 FC
All Others: 26W/m2 (2.4 W/sf)
Special:
--
Number of People:
Dressing & Toilet Rm: 1
Notes: 1. 600 mm x 600 mm (2' x 2') recessed
Waiting Area: 6
fluorescent light fixture, acrylic
Special Exhaust:
--
prismatic lens, W/2-FB31 lamps,
Notes:
--
3500 deg. K, CRI=75 (minimum).
PLUMBING AND MEDICAL GASES
POWER
Cold Water:
Yes
General:
Hot Water:
Yes
Emergency:
--
Laboratory Air:
--
Notes:
--
Laboratory Vacuum:
--
Sanitary Drain:
Yes
COMMUNICATION
Reagent Grade Water:
--
Medical Air:
--
ADP:
--
Medical Vacuum:
--
Radio:
--
Oxygen:
--
Telephone:
Yes
Notes:
--
Intercom:
--
Pub. Addr:
Yes
Notes: 1. Nurse call system in toilet room to
annunciate at reception area.
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 5-3