Final Draft 12-31-95
Revised May 1997
Equipment Guide List
CT Suite (Mid-Level Imaging System) - Gantry Room
SYMBOL
QTY AI
DESCRIPTION
NOTE: SERVICES, ELECTRICAL, AS REQUIRED FOR EQUIPMENT.
NOTE: LOW LEVEL DIMMED INCANDESCENT LIGHTING SYSTEM IN ADDITION TO GENERAL
NOTE: OUTLETS TO BE ON THE WALL, NEAR HEAD END OF X-RAY TABLE, 1524mm (60")
ABOVE THE FLOOR.
NOTE: OUTLET, WALL, MEDICAL AIR (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: OUTLET, WALL, OXYGEN (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: OUTLET, WALL, VACUUM (PG-18-1, MCS 15491; H-18-3, CS 310-1)
NOTE: SHIELDING, RADIATION, FOR ROOMS WITH
FIXED X-RAY EQUIPMENT, IN
ACCORDANCE WITH (PG-18-1, MCS 13091; H-18-3, CS 64-1; PG-18-4, SD 24 THRU 24K AND
NCRP REPORT NO. 33, 35 AND 49)
AR
CF
MODULAR CASEWORK
J-2
1
CF
SINK, CORROSION RESISTING STEEL, WITH CENTER DRAIN
OUTLET, 457mm x 381mm x 203mm (18" x 15" x 8")
(PG-18-1 & PG-18-6, MCS 11602)
1
VC
GANTRY AND MOUNTING FRAME
1
VC
TABLE, PATIENT AND MOUNTING FRAME
1
VV
CLOCK, BATTERY OPERATED
1
VV
DISPENSER, PAPER TOWEL, SURFACE MOUNTED
1
VV
DISPENSER, SOAP, LIQUID, WALL MOUNTED
1
VV
DISPOSAL, NEEDLES, WALL MOUNTED
1
VV
HOOK, COAT, WALL MOUNTED
1
VV
RACK, APRON, WALL MOUNTED
1
VV
RECEPTACLE, WASTE, STEP-ON TYPE APPROX 305mm (12")
DIAMETER
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 4-5