FINAL DRAFT 9-26-95
REVISED FEBRUARY 1997
Sleep Study Suite
Design Standards
Bedroom
ARCHITECTURAL
ADP:
---
Ceiling:
Acoustical ceiling tile
Radio:
---
Ceiling Ht:
2700 mm ( 9'-0")
Telephone:
Yes
Wall Finish:
Paint/ceramic tile
Intercom:
Microphone and outlets
Wainscot:
---
Public Addr:
---
Base:
Resilient base
Notes:
1. Provide closed circuit TV
Floor Finish:
cameras and receivers.
Slab Depr:
Slope bathroom floor
Notes:
---
CONDITIONING
SPECIAL EQUIPMENT
0
0
Dry Bulb Temp Cooling:
24 C (76 F)
None
0
0
25 C (78 F)
Minimum % Outside Air:
15%
ELECTRICAL
---
General:
30 FC
Noise Criteria:
NC-35
Notes:
1. Wall mounted fluorescent
Steam:
---
bed light. VA STD. detail
E-F-18.
Relative Humidity/Cooling:
50%
2. Wall mounted night light
Relative Humidity/Heating:
30%
fixture mounted 450 mm
(18") AFF.
Minimum Air Changes/Hr:
4
3. Provide electric for infra-red
lights.
Room Pressure:
0
4. Provide Faraday type
AC Load Equipment:
29 W/sm (2.7 W/sf)
shielding for bedrooms and
diagnostic rooms.
AC Loading Lighting:
11 W/sm (1.0 W/sf)
Number of People:
3
POWER
---
General:
Notes:
1. See HVAC Design Manual for
Emergency:
---
Hospital Projects for air balance
requirements between
Notes:
1. All power to bedrooms
bedroom
and the toilet /shower
will be controlled from
Diagnostic Rooms.
VA DESIGN GUIDE
PULMONARY MEDICINE SERVICE
PLATE 5-5