Final Draft 11-30-95
Revised March 1997
Design Standards
Prepacking & Extemporaneous Compounding
ARCHITECTURAL
CONDITIONING
Ceiling:
Ceiling Ht:
2740 mm (9'-0")
Dry Bulb Temp Cooling:
22C (72F)
Wall Finish:
Painted Gypsum Wallboard
22C (72F)
Wainscot:
--
Minimum % Outside Air:
15
Base:
Resilient Base
--
Noise Criteria:
NC-40
Slab Depr:
--
Steam:
--
Notes:
Relative Humidity/Cooling:
50%
Relative Humidity/Heating:
30%
SPECIAL EQUIPMENT
Minimum Air Changes/Hr.:
4
Room Pressure:
0
None
AC Load Lighting:
32 w/sm (3 w/sf)
LIGHTING
AC Load-Equipment:
54 w/sm (3 w/sf)
Number of People:
3
General:
100 FC
--
Special:
--
Notes:
Notes:
1. Lighting control by area;
switch(es) at room entrance.
PLUMBING AND MEDICAL GASES
2. 600mm x 1200mm (2' x 4')
Cold Water:
Yes
acrylic prismatic lens, W/3-F32T8
Hot Water:
Yes
lamps, 3500K, CRI=75
Laboratory Air:
--
(minimum)
Laboratory Vacuum:
--
Sanitary Drain:
Yes
POWER
Reagent Grade Water:
--
Medical Air:
--
General:
Medical Vacuum:
--
strip mold on modular casework).
Oxygen:
--
Emergency:
--
Notes: 1. Sanitary drain and vent need to be
Notes: 1.
Junction box for 120V power
chemical resistant type.
connection to electrical strip mold
on modular casework.
2.
Junction box for 120V power
connection to tablet and capsule
counting machine.
3.
Junction box for 120V power
connection to unit dose packer.
COMMUNICATION
ADP:
--
Radio:
--
Telephone:
Yes
Intercom:
--
Pub. Addr:
--
via modular case-work utility
access lab module to outlets on
casework.
VA DESIGN GUIDE
PHARMACY SERVICE
PAGE 5-4