FINAL DRAFT 9/6/95
UPDATED NOV. 1996
Equipment Guide List
Non-Imaging Procedure Room
SYMBOL
QTY AI
DESCRIPTION
NOTE: INTERCOM, STATION, HANDS FREE, WALL MOUNTED (PG-18-1, MCS 16760)
NOTE: OUTLETS, AIR, GAS AND VACUUM, GROUPED, 1220 mm (48") ON CENTER ABOVE
COUNTER (PG-18-1 & PG-18-6, MCS 11602)
NOTE: OUTLETS, EMERGENCY ALARM BUTTON , (PG-18-1 , MCS 16761)
NOTE: OUTLET, INTERCOM (EMPTY CONDUIT SYSTEM) (PG-18-1, MCS 16111)
NOTE: RECEPTACLE, ELECTRICAL, DUPLEX, 120 VOLT, 15 AMP, STRIP MOLD WITH
OUTLETS, WIRED ALTERNATELY ON SEPARATE CIRCUITS ON 610 mm (24") CENTERS
ABOVE COUNTER (PG-18-1, MCS 161403)
NOTE: RECEPTACLE, ELECTRICAL, DUPLEX, 120 VOLT, 20 AMP, (PG-18-1, MCS 16140;
NOTE:COUNTER TOPS OF MODULAR CASEWORK ARE TO BE IMPENETRABLE,
SEAMLESS SURFACES WITH RIMS AND DRIP GROOVES AT EDGES.
(PG-18-1, MCS 11620 PG-18-5)
AR
VV
910mm X 1220mm (36" X 48")
1
VV
CLOCK, BATTERY OPERATED
AR
CF
COUNTER TOP, CORROSION RESISTING STEEL,
32 mm (1-1/4") THICK
1
CC
DISPENSER, SOAP, WALL-MOUNTED WITH FOOT CONTROL
(PG-18-1, MCS 10800)
1
VV
DISPENSER, BI-FOLD PAPER TOWEL, SURFACE-MOUNTED
1
VV
DISPENSER, GLOVES, WALL-MOUNTED
AR
VV
REFRIGERATOR, 4 TO 10 DEGREES CELCIUS, 71 CUBIC M.
(25 CUBIC FT.), 120 VOLT, 20 AMP, REACH-IN TYPE, DOUBLE
DOOR, WITH SHELVES
1
VV
FREEZER, 4 TO 10 DEGREES CELCIUS, CAPACITY AS
REQUIRED,120 VOLT, 20 AMP, REACH-IN TYPE
H3-72
1
CC
FUME HOOD, RADIOISOTOPE, STAINLESS STEEL INTERIOR
WITH CABINET BASE, 2 EACH, WA-- AIR, GAS, VACUUM,
ELECTRIC DUPLEX OUTLETS AND CUP SINKS, BASE TO
SUPPORT SUPERSTRUCTURE PLUS 9.58 KPA (200 POUNDS
PER SQUARE FOOT), 1830mm X 910mm X 2130mm
(72" X 36" X 84") (PG-18-1 & PG-18-6, MCS 11610)
SEE HVAC DESIGN MANUAL FOR HOSPITAL PROJECTS.
AR
CF
MODULAR LABORATORY CASEWORK
1
VV
1
VV
RECEPTACLE, WASTE, BIOHAZARD TYPE, 762mm (30") DIA.,
ON WHEELS WITH FOOT-OPERATED LID
P-512
1
CC
SINK, CORROSION-RESISTING STEEL, SINGLE
COMPARTMENT WITH DRAINBOARD, WALL-HUNG 1,
MCS 15450)
VA DESIGN GUIDE
NUCLEAR MEDICINE SERVICE
5-3