JULY 2005
Design Standards
Gastroenterology Procedure / Treatment Room
Intercom:
--
ARCHITECTURAL
Nurse Call:
Yes
Coordinate with SFO; refer to Template SFO
Public Address:
--
Section 7, and Schedule C.
Radio/Entertainment:
--
Ceiling:
GWB Lay-in Panels
MATV:
--
Ceiling Height:
2740 mm (9'-0")
CCTV:
--
Wall Finish:
Painted GWB
MID:
--
Wainscot:
--
Security/Duress:
--
Base:
Resilient Sheet Flooring
VTEL:
--
Floor Finish:
Resilient Sheet Flooring
VA Satellite TV:
--
Slab Depression:
--
Notes:
--
Sound Protection:
STC 40
Notes:
--
CONDITIONING
LIGHTING
Coordinate with SFO; refer to Template SFO
Coordinate with SFO; refer to Template SFO
Section 6.5.
Section 6.7, 6.7.1, 6.7.5.
22C (73F)
Dry Bulb Temp Cooling:
General:
50-100 FC
26C (78F)
Special:
--
Minimum % Outside Air:
Notes:
Yes
cent light fixture with
Noise Criteria:
NC-40
acrylic prismatic lens.
Steam:
--
2. 50% of light fixture on
Relative Humidity/Cooling:
50%
emergency system
Relative Humidity/Heating:
30%
critical power source.
Minimum Air Changes/Hr.:
10
Room Pressure:
Negative
POWER
AC Load-Equipment:
6 W/SF
Coordinate with SFO; refer to Template SFO
AC Load-Lighting:
1/6 W/SF
Section 6.7, 6.7.1, 6.7.3, 6.7.4 and Schedule B.
Number of People:
3
General:
Yes
--
Emergency:
Yes
Notes:
--
Notes:
1. Junction box x-ray
illuminator. Connect
PLUMBING AND MEDICAL GASES
per equipment
Coordinate with SFO; refer to Template SFO
manufacturer.
Section 6.6 and Schedule B
mounted junction box
Cold Water:
Yes
for exam and treatment
Hot Water:
Yes
table. Connection to
Laboratory Air:
--
equipment per manu-
Laboratory Vacuum:
--
facturer. Provide
Sanitary Drain:
Yes
dedicated circuit. Verify
Reagent grade Water:
--
wattage, voltage, and
Medical Air:
--
phase.
Medical Vacuum:
--
3. Portable exam light.
Oxygen:
--
Notes:
--
COMMUNICATION/SPECIAL SYSTEMS
Coordinate with SFO; refer to Template SFO
Sections 6.7.8, 6.7.9, and Schedule B.
Data:
Yes
Telephone:
Yes
VA DESIGN GUIDE
APPENDIX B
LEASE BASED OUTPATIENT CLINICS
3-8.3