SECTION 11620
LABORATORY CUSTOM FABRICATED EQUIPMENT
1.
This section provides inspection and test procedures for Laboratory
custom fabricated equipment including animal necropsy dissecting tables,
necropsy sinks, cage scullery sinks, and animal feeder bottles filling
stations.
2.
Symbols
Item Description
HW-800 ..... Necropsy Dissecting Table with down draft exhaust.
HW-801 ..... Wall mounted Necropsy Dissecting Table with mobile table.
HW-802 ..... Wall mounted Necropsy Dissecting Table.
HW-803 ..... Floor mounted Necropsy Dissecting Table.
HW-815 ..... Wall mounted Necropsy Sink.
HW-879 ..... Animal cage scullery sink and table with disposer.
HW-900 ..... Feeder bottle filling station.
3.
INSPECTIONS AND TESTS
A.
Visual Examination:
Each table, sink and accessories should be examined
for the following defects.
1)
Types, class and size is not as specified.
2)
Markings illegible or missing.
3)
Damaged or defects affecting function or serviceability.
4)
Dimensions not as specified, or not within allowable tolerances.
5)
Metal thickness not as specified.
6)
Edges and surfaces not finished smooth.
7)
Finish not as specified.
8)
Finished surfaces not free of discoloration or stains.
9)
Parts or components missing.
10)
Any component fractured, buckled, bent, punctured, malformed or
damaged.
11)
Parts not joined or connected as specified.
12)
Threaded fasteners broken, stripped, or fractured.
13)
Welding not type specified, incomplete, burn holes, cracked, or
fractured.
14)
Equipment, including fittings is not designed to prevent the
backflow of polluted water or waste into water supply system.
15)
Faucets and other service fixtures not type specified.
16)
Leaks at faucets, traps, sinks , and other accessories.
17)
Sinks and tables not installed leveled.
18)
Backsplashes height is not as specified.
19)
Backsplashes not sealed to wall or adjoining surfaces as
specified.
20)
When specified, disposer does not operate as specified.
B.
Certification Compliance Examination:
Certifications, certified test reports,
or listing marks for codes and standards, as applicable, should be
examined and validated as proof of compliance. Conformance should be
evidence by seal, label, stamp or certified test reports.