FINAL DRAFT 10-31-95
REVISED January 1997
Design Standards
Phototherapy TreatmenRm/Phototherapy Shower m
t
R
ARCHITECTURAL Treatment Room
-
COMMUNICATION
Ceiling:
ADP:
--
Ceiling Ht:
2700 mm (9'-0")
Radio:
--
Wall Finish:
Paint
Telephone:
Yes
Wainscot:
--
Intercom:
--
Base:
Resilient sheet flooring integral
Pub. Addr:
--
base
Notes:
1. Nurse call in patient shower to
Floor Finish: Resilient
be mounted 1200 mm (48")
Slab Depr:
--
AFF.
Notes:
1. 900 mm (3'-0") wide door.
ARCHITECTURAL Shower/Toilet
-
CONDITIONING
Ceiling:
Portland cement plaster
Dry Bulb Temp Cooling:
24C (76F)
Ceiling Ht:
2700 mm (9'-0")
25C (78F)
Wall Finish:
Ceramic tile, full height
Minimum % Outside Air:
100
Wainscot:
--
Yes
Base:
Ceramic tile
Noise Criteria:
NC-40
Floor Finish: Ceramic tile
Steam:
--
Slab Depr:
75 mm (3") shower depression
Relative Humidity/Cooling:
50%
Notes:
1. 900 mm (3'-0") wide door.
Relative Humidity/Heating:
30%
Minimum Air Changes/Hr.:
8
SPECIAL EQUIPMENT
Room Pressure:
Negative
2
AC Load Lighting:
46 W/m (4.2 W/sf)
None
AC Load-Equipment:
(5 W/ sf to10 W/sf)
LIGHTING
54 W/sm - 109 W/sm
Number of People:
3
General:
Treat. Room 50 FC
Yes
Toilet/shower 30 FC
Notes: 1. Coordinate equipment heat gain
Special:
--
requirements with supplier.
Notes:
1. 600 mm x 1200 mm (2'x4')
recessed fluorescent light
PLUMBING AND MEDICAL GASES
fixture, acrylic prismatic lens,
W/3-F32T8 lamps, 3500K,
Cold Water
Yes
Hot Water:
Yes
CRI=75 (minimum).
Laboratory Air:
--
Laboratory Vacuum:
--
fluorescent light fixture, acrylic
Sanitary Drain:
Yes
prismatic lens, W/2-FB31
lamps, 3500K, CRI=75
Reagent Grade Water:
--
Medical Air:
--
(minimum).
Medical Vacuum:
--
Oxygen:
--
light, drop opal lens, 150W-A21
Notes:
lamp, UL listed for wet
locations.
POWER
General:
Emergency:
--
Notes:
1. Junction box for 208V power
connection to UV light boxes.
VA DESIGN GUIDE
AMBULATORY CARE (Hospital Based)
PAGE 6-12