Design Standards
Gastric (Esophageal) Motility Procedure Room
ARCHITECTURAL
CONDITIONING
Ceiling:
Ceiling Ht:
2700 mm (9'-0")
Dry Bulb Temp Cooling:
25C (78F)
Wall Finish:
Paint
22C (72F)
Ceramic Tile in Toilet Rm.
Minimum % Outside Air:
--
Wainscot:
--
Yes
Base:
Resilient sheet flooring int egral
Noise Criteria:
NC-40
base/Ceramic Tile in Toilet Rm.
Steam:
--
Floor Finish: Resilient sheet flooring
Relative Humidity/Cooling:
50%
Ceramic Tile in Toilet Rm.
Relative Humidity/Heating:
30%
Slab Depr:
--
Minimum Air Changes/Hr.:
8
Notes:
1. 900 mm (3'-0") wide door.
Room Pressure:
Negative
2
AC Load Lighting:
19 W/m (1.8 W/sf)
65 W/m 2 (6 W/sf)
SPECIAL EQUIPMENT
AC Load-Equipment:
Number of People:
3
None
Yes
LIGHTING
Notes:
General:
50 FC ambient/100 FC task
PLUMBING AND MEDICAL GASES
Special:
--
Notes:
1. Dual-switching provides 50/100
Cold Water:
Yes
FC levels.
Hot Water:
Yes
2. 600 mm x 1200 mm (2'x4')
Laboratory Air:
--
Laboratory Vacuum:
--
fixture, acrylic prismatic lens,
Sanitary Drain:
Yes
W/3-F32T8 lamps, 3500K,
Reagent Grade Water:
--
Medical Air:
--
CRI=75 (minimum).
Medical Vacuum:
--
3. 600 mm x 600 mm (2'x2')
Oxygen:
--
Notes:
fixture, acrylic prismatic lens,
W/2-FB31 lamps, 3500 K
CRI=75 (minimum)
POWER
General:
(dedicated recept./circuit for
exam table).
Emergency:
--
Notes:
COMMUNICATION
ADP:
Yes
Radio:
--
Telephone:
Yes
Intercom:
--
Pub. Addr:
--
Notes
VA DESIGN GUIDE
AMBULATORY CARE (Hospital Based)
PAGE 6-9