Design Standards
ScreeningProctoscopy/Sigmoidoscopy Room
ARCHITECTURAL
CONDITIONING
Ceiling:
Gyp Wall Bd in Toilet Rm.
Dry Bulb Temp Cooling:
25C (78F)
Ceiling Ht:
2700 MM (9'-0")
22C (72F)
Wall Finish:
Paint
Minimum % Outside Air:
--
Ceramic Tile in Toilet Rm.
Yes
Wainscot:
--
Noise Criteria:
NC-40
Base:
Resilient sheet flooring integral
Steam:
--
base
Relative Humidity/Cooling:
50%
Floor Finish: Resilient sheet flooring
Relative Humidity/Heating:
30%
Ceramic Tile in Toilet Rm.
Minimum Air Changes/Hr.:
6
Slab Depr:
--
Room Pressure:
Negative
2
Notes:
1. 1200 MM (4'-0") wide doors.
AC Load Lighting:
30 W/m (2.8 W/sf)
54 W/m 2 (5W/sf)
AC Load-Equipment:
SPECIAL EQUIPMENT
Number of People:
4
--
None
Notes:
LIGHTING
PLUMBING AND MEDICAL GASES
General:
Procto Rm:100 FC
Toilet: 30 FC
Cold Water:
Yes
Special:
--
Hot Water:
Yes
Notes:
1. 50% of lighting on critical
Laboratory Air:
--
power.
Laboratory Vacuum:
--
2. 600 mm x 1200 mm (2'x4')
Sanitary Drain:
Yes
Reagent Grade Water:
--
fixture, acrylic prismatic lens,
Medical Air:
Yes
W/3-F32T8 lamps, 3500K,
Medical Vacuum:
Yes
Oxygen:
--
CRI=75 (Minimum)
Notes:
1. Potential cold water is required
3. 600 mm x 600 mm (2'x2')
for laser equipment. (Coordi-
nate with equipment supplier.)
fixture, acrylic prismatic lens,
W/2-FB31 lamps, 3500K,
CRI=75 (minimum).
POWER
General:
Emergency:
200W (receptacle)
Notes:
1. Special receptacle for laser
equipment. (Coordinate with
supplier.)
COMMUNICATION
ADP:
Yes
Radio:
--
Telephone:
Yes
Intercom:
--
Pub. Addr:
--
Notes:
1. Nurse call
VA DESIGN GUIDE
AMBULATORY CARE (Hospital Based)
PAGE 6-8