First Draft 3-15-97
Design Standards
Gastric Motility Studies
ARCHITECTURAL
CONDITIONING
Ceiling:
Dry Bulb Temp Cooling:
25C (78F)
Ceiling Ht:
2700 MM (9'-0")
22C (72F)
Wall Finish:
Paint
Minimum % Outside Air:
--
Ceramic Tile in Toilet Room
Yes
Wainscot:
--
Noise Criteria:
NC-40
Base: Resilient sheet flooring w/ integral base
Steam:
--
Ceramic Tile in Toilet Room
Relative Humidity/Cooling:
50%
Floor Finish:
Resilient sheet flooring
Relative Humidity/Heating:
30%
Ceramic Tile in Toilet Room
Minimum Air Changes/Hr.:
8
Slab Depr:
--
Room Pressure:
Negative
Notes: 1. 900mm (3'-0") wide entry door.
AC Load Lighting:
19 w/sm (1.8 w/sf)
SPECIAL EQUIPMENT
AC Load-Equipment:
65 w/sm (6 w/sf)
None
Number of People:
3
Yes
LIGHTING
Notes:
--
General:
50 FC (ambient)/100 FC Task
PLUMBING AND MEDICAL GASES
Special:
--
Cold Water:
Yes
Notes:
1. Dual-level switching provided
Hot Water:
Yes
50/10 FC levels.
Laboratory Air:
--
fluorescent light fixture, acrylic
Laboratory Vacuum:
--
prismatic lens, W/2-F32T8
Sanitary Drain:
Yes
lamps, 3500K, CRI=75 (Min.).
Reagent Grade Water:
--
Medical Air:
--
3. 600mm x 600mm (2' x 2')
Medical Vacuum:
--
Oxygen:
--
fixture, acrylic prismatic lens,
Notes:
--
W/3-FB31 lamps, 3500K,
CRI=75 (minimum)
POWER
General:
1000W (dedicated recept./circuit for exam table)
Emergency:
--
Notes:
--
COMMUNICATION
ADP:
Yes
Radio:
--
Telephone:
Yes
Intercom:
--
Pub. Addr:
--
Notes:
--
VA DESIGN GUIDE
AMBULATORY CARE INVASIVE PROCEDURES SUITE
PAGE 5-5