SYMBOL QTY AI
DESCRIPTION
MCS
WINDOW, VIEWING, FOR PATIENT OBSERVATION
AR CC
13091
NOTE: CONSTRUCTION TO COMPLY WITH NFPA 101.
RECEPTACLE MODULES; ONE MODULE ON EACH WALL, EACH MODULE SHALL CONSIST
4
CC OF FOUR SINGLE, 120V, 20 AMP, HOSPITAL GRADE TYPE RECEPTACLES, EACH ON
16140
EMERGENCY POWER & ON A SEPARATE CIRCUIT.
ILLUMINATOR, FILM, X-RAY, RECESSED, 120 VOLT, 20 AMP, 14" X 17" (355 mm X 430 mm)
AR CC
16510
(INSTALLATION NOT TO BE COMBINED WITH IPU'S OR OTHER ELECTRICAL DEVICES)
16510
COLUMN, ARTICULATING OR TELESCOPING, CEILING MOUNTED COLUMN A: LOCATE AT
HEAD OF TABLE, 48" (1220 mm) TO 72" (1830 mm) FROM THE CENTERLINE AND 24" (610
mm) TO 48" (1220 mm) TO THE LEFT OF THE CENTERLINE OF TABLE. COLUMN B:
AR CC LOCATE AT THE FOOT OF THE TABLE, 48" (1220 mm) TO 72" (1830 mm) FROM THE
15491
CENTERLINE AND 24" (610 mm) TO 48" (1220 mm) TO THE RIGHT OF THE CENTERLINE OF
THE TABLE. EACH COLUMN CONTAINS THE FOLLOWING:
2
INLET, VACUUM
15491
1
OUTLET, NITROUS OXIDE
15491
2
OUTLET, OXYGEN
15491
1
OUTLET, MEDICAL AIR
15491
1
15491
1
INLET, VACUUM, DEDICATED ANESTHESIA GAS EVACUATION
15491
INLET, MASS SPECTROMETER (BLANK OUTLET) DATA CONNECTION
1
TELECOMMUNICATIONS CONNECTION 4 SINGLE, 120V, 20 AMP HOSPITAL GRADE
16140
AR
16140
Design Guide - Surgical Service
Equipment Guide List
Special Purpose Operating Rm
Guide Plate:
4-2j (i)
Page 1 of 4
MAX: 800 NSF (74.40 NSM)
MIN: 600 NSF (55.80 NSM)
MIN: 700 NSF (65.10 NSM) CARDIAC
Date: August 2005