MCS
SYMBOL QTY AI DESCRIPTION
AR CC WINDOW, VIEWING, FOR PATIENT OBSERVATION
13091
RECEPTACLE MODULES; ONE MODULE ON EACH WALL, EACH MODULE SHALL CONSIST OF
16140
AR CC THREE SINGLE, 120V, 20 AMP, HOSPITAL GRADE TYPE RECEPTACLES, EACH ON
EMERGENCY POWER & ON SEPARATE CIRCUIT.
ILLUMINATOR, FILM, X-RAY, RECESSED, 120 VOLT, 20 AMP, 14" X 17" (355 mm X 430 mm)
AR CC
16510
(INSTALLATION NOT TO BE COMBINED WITH IPU'S OR OTHER ELECTRICAL DEVICES)
16510
COLUMN, ARTICULATING OR TELESCOPING, CEILING MOUNTED COLUMN A: LOCATE AT
HEAD OF TABLE, 48" (1220 mm) TO 72" (1830 mm) FROM THE CENTERLINE AND 24" (610
mm) TO 48" (1220 mm) TO THE LEFT OF THE CENTERLINE OF TABLE. COLUMN B: LOCATE
15491
AR CC AT THE FOOT OF THE TABLE, 48" (1220 mm) TO 72" (1830 mm) FROM THE CENTERLINE
AND 24" (610 mm) TO 48" (1220 mm) TO THE RIGHT OF THE CENTERLINE OF THE TABLE
EACH COLUMN CONTAINS THE FOLLOWING (EXACT TYPE AND QUANTITY SHALL BE
COORDINATED WITH VAMC).
15491
INLET, VACUUM
AR
OUTLET, NITROUS OXIDE
AR
15491
15491
OUTLET, OXYGEN
AR
15491
OUTLET, MEDICAL AIR
AR
15491
AR
INLET, VACUUM, DEDICATED ANESTHESIA GAS EVACUATION
AR
15491
INLET, MASS SPECTROMETER (BLANK OUTLET) DATA CONNECTION
TELECOMMUNICATIONS CONNECTION 4 SINGLE, 120V, 20 AMP HOSPITAL
AR
16140
ISOLATED POWER UNIT PROVIDES ISOLATED ELECTRICAL POWER, INCLUDES LINE
CC
2
16675
ISOLATION MONITOR, ISOLATION TRANSFORMER AND CIRCUIT BREAKERS
LIGHT, MAJOR, SURGICAL WITH VARIABLE INTENSITY CONTROL, SUSPENSION AS
16515
AR CF
REQUIRED, CEILING MOUNTED
Design Guide - Surgical Service
Equipment Guide List
General Operating Room
Guide Plate:
4-1j (i)
Page 1 of 3
MAX: 650 NSF (60.45 NSM)
MIN: 450 NSF (41.86 NSM)
Date: August 2005