Final Draft 12-31-95
Revised May 1997
Design Standards
Dedicated Chest Room
ARCHITECTURAL
POWER (cont.)
Notes: 3. 480V., 3P., 150A., adjustable trip flush
Ceiling:
mounted circuit breaker with shunt
Ceiling Ht:
2890 mm (9'-6")
trip. Run empty 50 mm (2" C.) from
Wall Finish:
Paint
circuit breaker to x-ray duct above
Wainscot:
--
finished ceiling.
Base:
Resilient base
4. 240V, 2P., 30A flush MTD. circuit
Floor Finish:
breaker for film processor.
Slab Depr:
100 mm (4")
5. 250 mm x 140 mm (10' x 5-") deep
Notes:
1. 1200 mm (4'-0") wide door. See
flush vertical wall duct with 300 mm
VA Construction Standard 64-1,X-
(12") wide screw-on cover. Connect to
x-ray duct above finished ceiling and
Special Control".
terminate at finished floor.
6. 250 mm x 140 mm (10' x 5-") deep
SPECIAL EQUIPMENT
x-ray duct above finished ceiling with
None
250 mm (10") wide screw-on cover.
Connect to vertical wall duct.
LIGHTING
7. Emergency Power Off (EPO) push
General:
30 FC
button station. Connect to shunt trip
Special:
--
circuit breakers.
Notes:
1. Continuous fluorescent
direct/indirect wall mounted light
fixture, acrylic prismatic lens on top
COMMUNICATION
and bottom, W/2-F32T8 lamps,
ADP:
--
3500 deg. K, CRI=75 (min.)
Radio:
--
Telephone:
Yes
downlight with 100W, A-21 lamp.
Intercom:
--
Pub. Addr:
--
controlled by dimmer.
Notes:
1.Nurse Call
POWER
General:
Emergency:
--
Notes: 1. The electrical plan as shown is to
be used as a guide only.
Equipment locations, dimensions
and wiring requirements to be per
the x-ray system suppliers
equipment installation drawings.
2. Electrical trades shall provide
necessary conduit, openings,
bushings, nipples, flexible conduit
etc., as required at the various
junction boxes, duct and conduit
terminations to allow proper
connections of the x-ray equipment
and related accessories.
VA DESIGN GUIDE
RADIOLOGY SERVICE
PAGE 4-2