FINAL DRAFT 9/6/95
UPDATED NOV. 1996
Design Standards
Radioimmunoassay Unit
ARCHITECTURAL
CONDITIONING
Ceiling:
Acoustical ceiling tile
0
0
Ceiling Ht:
2700 mm (9'-0")
Dry Bulb Temp Cooling:
25 C (78 F)
0
0
Wall Finish:
Paint
22 C (72 F)
Wainscot:
---
Minimum % Outside Air:
--
Base:
Integral welded seam
Yes
sheet flooring base
Noise Criteria:
NC-40
Floor Finish:
Welded seam sheet
Steam:
---
flooring
Relative Humidity/Cooling:
50%
Slab Depr.:
---
Relative Humidity/Heating:
30%
Notes:
1220 mm (4'-0") wide doors
Minimum Air Changes/Hr.:
12
Room Pressure:
Negative
SPECIAL EQUIPMENT
AC Load-Equipment:
195 w/sm (18 w/sf)
AC Load Lighting:
22.5 w/sm (2.1 w/sf)
---
Number of People:
12
ELECTRICAL LIGHTING
Provide safety cabinet
General:
50FC
exhaust
Notes:
1.
Lighting control by area;
Notes
1. Verify equipment AC load as
switch(ex) at room entrance.
per actual equipment used.
2. 600 mm x 1200 mm (2'
x 4')
PLUMBING AND MEDICAL GAS
fluorescent
light
fixtures, acrylic prismatic
lens,
Cold Water:
Yes
W/2-=F32T8 lamps, 4100
deg.
Hot Water:
Yes
K, CRI-85 (minimum).
Laboratory Air:
Yes
Laboratory Vacuum:
Yes
POWER
Sanitary Drain:
Yes
Reagent Grade Water:
Yes
General:
Medical Air:
----
strip mold on modular casework)
Medical Vacuum:
----
Emergency:
Oxygen:
----
Notes:
1. Junction
box
for
120V
Notes:
1. Waste and vent needs to be
power/connection to electrical
chemical resistant type.
strip/mold
on
modular
casework.
2. Natural gas is required for
biohazard safety cabinet.
safety cabinet on emergency
power.
ADP:
Yes
Radio:
---
Telephone:
Yes
Intercom:
Yes
Public Addr:
---
Notes:
via modular casework utility
access lab module to outlets
on casework.
VA DESIGN GUIDE
NUCLEAR MEDICINE SERVICE
5-4