EQUIPMENT GUIDE LIST - CHAPTER 276
07-02
RADIOLOGY SERVICE
SYMBOL
QNTY
ACQ/INS
DESCRIPTION
SPEC
REFERENCE
DETAIL
1
VV
HOOK, COAT, WALL MOUNTED
1
VV
RACK, APRON, WALL MOUNTED
1
VV
DISPOSAL, NEEDLES, WALL MOUNTED
CONTROL AREA:
AR
CC
WINDOW, LEAD GLASS, MINIMUM 457 mm X 610 mm (18" X 24")
13091
CONSOLE, CONTROL, X-RAY, FOR RADIOGRAPHIC UNIT (CAPACITY
1
VC
OF GENERATOR TO BE DETERMINED ON PROJECT BASIS)
1
CC
INTERCOM, STATION
16760
1
CC
OUTLET, INTERCOM (EMPTY CONDUIT SYSTEM)
16111
NOTE: PROVIDE INTERCOM STATION IN CONTROL BOOTH
CONNECTED TO REMOTE STATION IN X-RAY ROOM, WHEN CONTROL
BOOTH IS SEPARATED FROM X-RAY ROOM.
AR
CC
LIGHT, WARNING (DO NOT ENTER) OVER DOOR
FILM PROCESSING AREA:
PROCESSOR, FILM, AUTOMATIC, THRU-THE-WALL TYPE, CAPACITY,
1
VC
120 FILMS PER HOUR, 813 mm X 889 mm X 1219 mm (32" X 35" X 48")
-OR-
PROCESSOR, FILM, AUTOMATIC, THRU-THE-WALL TYPE, CAPACITY,
1
VC
185 FILMS PER HOUR, 762 mm X 2184 mm X 1118 mm (30" X 86" X 44")
-OR-
PROCESSOR, FILM, AUTOMATIC, 90 SECOND PROCESSING CYCLE,
1
VC
THRU WALL
FLOOR DRAIN, NEAR FILM PROCESSING UNITS (PLUMBING DESIGN
1
CC
CRITERIA AND INSTRUCTIONS)
TANK, REPLENISHER, 95 L (25 GAL.) CAPACITY AND CURBING, 2 FOR
1
VC
EACH PROCESSOR
FLOOR DRAIN, NEAR REPLENISHER TANK UNITS (PLUMBING DESIGN
1
CC
CRITERIA & INSTRUCTIONS)
HOSE BIBB, COMBINATION FAUCET, WALL MOUNTED, CONCEALED
P-802
1
CC
SUPPLY PIPES
15450
SINK, CORROSION RESISTING STEEL, SINGLE COMPARTMENT,
P-516
1
CC
WALL HUNG
15450
Page 32