b.
Slab depression is not required for CTR. Do not use
ramps at doors.
c.
Provide separate room with floor drain to accommodate
the water cooled compressor. Compressor Room should be
no more than 4.5 m (15 linear feet) from the
applicable CTR.
d.
Show the CTR control panel location.
e.
Show isometric view of CTR as shown in PG-18-6.
f.
Equipment must comply with the requirements stated in
Specification Section 11615.
3.
Custom Fabricated Laboratory Equipment
a.
Symbols for this equipment are prefixed by "HW".
b.
Completely detail each custom fabricated item.
c.
Provide adequate clearances.
d.
Equipment must comply with the requirements stated in
Specification Section 11620.
4.
Solution Warming Cabinets
a.
Symbols for this equipment are prefixed by "HW".
b.
Show elevation detail as shown in PG-18-6.
c.
Provide adequate loading and servicing clearances in
compliance with PG-18-6.
d.
Equipment must comply with the requirements stated in
Specification Section 11701.
5.
Medical Washing Equipment
a.
Symbols for this equipment is prefixed by "HW".
b.
Recess equipment into mechanical spaces where possible
utilizing metal modular walls paneling.
c.
Provide adequate loading and servicing clearances in
compliance with PG-18-6
d.
Provide reagent grade III or better water for final
rinse to glassware washers that are required to have a
mineral free water cycle. Coordinate with plumbing
drawings and specifications.
e.
Provide details of pit for cart washers. Coordinate
with the structural drawings and specifications.
f.
Provide elevation drawings of all Medical Washing
Equipment as shown in PG-18-6.
g.
Equipment must comply with the requirements stated in
Specification Section 11714.
6.
Morgue Equipment
5