FINAL DRAFT 10-31-95
REVISED January 1997
Design Standards
AerosolizedPentamidine Procedure Room
ARCHITECTURAL
CONDITIONING
Ceiling:
Gypsum wallboard lay-in panels
Ceiling Ht:
2700 mm (9'-0")
Dry Bulb Temp Cooling:
24C (76F)
Wall Finish:
Paint.
25C (78F)
Wainscot:
--
Minimum % Outside Air:
--
Base:
Resilient base
Yes
Noise Criteria:
NC-35
Slab Depr:
--
Steam:
--
Notes:
1. Provide STC 40 partitions and
Relative Humidity/Cooling:
50%
door.
Relative Humidity/Heating:
30%
2. 900 mm (3'-0") wide door.
Minimum Air Changes/Hr.:
10
Room Pressure:
Negative
2
SPECIAL EQUIPMENT
AC Load Lighting:
13 W/m (1.2 W/sf)
32 W/m 2 (3 W/sf)
AC Load-Equipment
None
Number of People:
Provide 3
Special Exhaust:
LIGHTING
Notes:
General:
50 FC
Special:
--
PLUMBING AND MEDICAL GASES
Notes:
1. 600 mm x 1200 mm (2'x4')
Cold Water:
Yes
fixture, acrylic prismatic lens,
Hot Water:
Yes
W/2-F32T8 lamps, 3500K,
Laboratory Air:
--
Laboratory Vacuum:
--
CRI=75 (minimum).
Sanitary Drain:
Yes
Reagent Grade Water:
--
POWER
Medical Air:
Yes
General:
Medical Vacuum:
--
Emergency:
--
Oxygen:
--
Notes:
Notes:
COMMUNICATION
ADP:
--
Radio:
--
Telephone:
Yes
Intercom:
--
Pub. Addr:
--
Notes:
VA DESIGN GUIDE
AMBULATORY CARE (Hospital Based)
PAGE 6-16