HVAC REQUIREMENTS IN RADIATION THERAPY
(REFER TO VA DESIGN GUIDE, RADIATION THERAPY, FOR ADDITIONAL INFORMATION)
1. Indoor Design Conditions:
Area
Summer
Winter
HVAC Design
Degree C
Degree C
Manual Ref.
(Degree F)
(Degree F)
Radiation Therapy
24(76) 50% RH
25(78) 30% RH
Table 1-1
(Linear Accelerators)
2. Maintain negative pressure in Radiation Therapy
Table 1-4
3. See VA Design Guide for:
Treatment Planning,
Superficial Therapy Unit,
OrthovoltageTherapy Unit,
SupervoltageTherapy Unit,
Megavoltage Therapy Unit (10MeV)
Mold Room, and
Treatment Planning and Computer Room
4. Maintain NC-35 to 40 depending upon use of the space
1.3.12
5. Filtration: Pre-filters 30%, and After-filters 85%
Table 2-1
2.11.3
hoods.
7. Emergency power requirements for hoods and exhaust fans, and
2.16
closed loop water chilling equipment for linear accelerators.
8. Lead wrapping of duct work penetrating lead lined walls.
3.19.2
9. Requirements for Linear Accelerators
3.15