FINAL DRAFT 2-29-96
Revised 10/96
Equipment Guide List
Superficial Therapy Unit
SYMBOL
QTY AI
DESCRIPTION
NOTE: THE FOLLOWING X-RAY EQUIPMENT TO BE DETERMINED ON PROJECT
BASIS BY VETERANS HEALTH ADMINISTRATION.
1
VC
1
VC
POWER UNIT DEVICE
2
VC
COOLER, WATER
1
VC
STABILIZER
INSTALL POWER UNIT DEVICE, WATER COOLER AND
STABILIZER IN AN ENCLOSED SPACE NOT LESS THAN,
154mm x 3048mm (5'-0" x 10'-0") WITH VENTILATION TO
DISSIPATE HEAT GENERATED, AMOUNT OF
EQUIPMENT VARIES WITH MANUFACTURER.
NOTE: SERVICES, ELECTRICAL & PLUMBING, AS
REQUIRED FOR EQUIPMENT.
1
VV
TABLE, THERAPY, HYDRAULIC, MOBILE
1
CC
DOOR, POWER OPERATED, WITH ELECTRIC SAFETY
INTERLOCK TIED INTO CONTROL PANEL WITH LEAD
GLASS WINDOW (PG-18-1, MCS 13091; PG18-4, SD 24G)
1
CC
WINDOW, VIEWING, LEAD GLASS, FOR PATIENT
OBSERVATION (PG-18-1, MCS 13091; PG-18-4, SD 25G)
AR
CC
SHIELDING, RADIATION, FOR ROOMS WITH FIXED X-RAY
EQUIPMENT, IN ACCORDANCE WITH (PG-18-1, MSC 13091,
H-18-3, CS 64-1; PG-18-4, SD 24 THRU 24K AND NCRP
REPORT NO. 33, 35 AND 49)
1
CC
SPEAKER, WITH CHANNEL SELECTOR AND VOLUME
CONTROL, HOSPITAL RADIO SYSTEM (PG-18-1, MCS 16771)
VA DESIGN GUIDE
GUIDE PLATE 4-2