FINAL DRAFT 11/7/96
Equipment Guide List
SYMBOL
QTY AI
DESCRIPTION
General Notes
NOTE: ALL MATERIALS INSTALLED, LOCATED WITHIN OR INTENDED TO ENTER THE
GANTRY AREA, MUST BE NON-MAGNETIC EXCEPT FOR RECOMMENDED MAGNET
SHIELDING MATERIALS ON OR IN WALLS, FLOORS OR CEILINGS.
NOTE: ALL PENETRATIONS (THROUGH THE RF SHIELDED ROOM FOR PLUMBING,
ELECTRICAL, MECHANICAL) OR ANY CONDUCTIVE BUILDING MATERIAL SHALL BE
ELECTRICALLY ISOLATED AND RFI FILTERED IN ACCORDANCE WITH
MANUFACTURERS STANDARDS.
NOTE: CONDUCTIVE CONNECTIONS ARE TO BE MADE OUTSIDE THE RF ROOM.
NOTE: PROVIDE NON-MAGNETIC MATERIAL FOR AIR, OXYGEN AND VACUUM
OUTLETS FROM PORTABLE SUPPLY, ONLY WHEN APPROVED FOR A SPECIFIC
PROJECT BY VHA.
Gantry Room
NOTE: INTERCOM STATION, CONNECTED TO STATION IN THE CONTROL ROOM.
NOTE: INTERCOM, STATION, NON-MAGNETIC MATERIAL (PG-18-1, MCS 16760
NOTE: COMPUTER/ACCESS FLOORING, NON-MAGNETIC MATERIALS)
NOTE: CONNECTIONS, PLUMBING, ELECTRICAL OR MECHANICAL AS REQUIRED
NOTE: LIGHTING SYSTEM, DC, FOR INCANDESCENT LIGHT FIXTURES,
NON-MAGNETIC MATERIAL, MULTI-CIRCUIT LIGHT FIXTURES
NOTE: OUTLET, INTERCOM (EMPTY CONDUIT SYSTEM) (PG 18-1, MCS 16111)
NOTE: RADIO FREQUENCY COPPER GROUND PLANES, LOCATED ON
THE FINISHED
FLOOR SURFACE, FROM THE MAGNET TO THE ELECTRONIC CABINETS
NOTE: SHIELDING, MAGNETIC (ACTIVE OR PASSIVE)
NOTE: SYSTEM EMERGENCY OFF BUTTON PANEL, GUARD COLLAR, FLUSH
MOUNTED, LOCATED NEAR EXIT FROM GANTRY ROOM, APPROX., 1524 MM (60")
ABOVE FINISHED FLOOR
AR
CC
MODULAR CASEWORK
AR
VC
GANTRY, MAGNET, PATIENT COUCH, SIZES AS REQUIRED
1
VC
OXYGEN MONITOR AND ALARM, WALL MOUNTED,
APPROX., 279 MM X 178 MM X 229 MM (11" X 7" X 9")
LOCATED APPROX., 1524 MM (60") ABOVE FINISHED FLOOR
VA DESIGN GUIDE
4-1