Design Standards
Radiology Storage, Unexposed Film & Chemicals
Security/Duress:
--
ARCHITECTURAL
VTEL:
--
Coordinate with SFO; refer to Template SFO
VA Satellite TV:
--
Section 7, and Schedule C.
Notes:
--
Ceiling:
GWB
Ceiling Height:
2740 mm (9'-0")
Wall Finish:
Painted GWB
CONDITIONING
Wainscot:
--
Coordinate with SFO; refer to Template SFO
Base:
LM
Section 6.5.
Floor Finish:
LM
26C (78F)
Dry Bulb Temp Cooling:
Slab Depression:
--
22C (72F)
Sound Protection:
--
Minimum % Outside Air:
Notes:
--
Yes
Noise Criteria:
NC-40
LIGHTING
Steam:
--
Coordinate with SFO; refer to Template SFO
Relative Humidity/Cooling:
50%
Section 6.7, 6.7.1, 6.7.5.
Relative Humidity/Heating:
30%
Minimum Air Changes/Hr.:
4
General:
30 FC
Room Pressure:
Negative
Special:
1. Provide 1' X 4' surface
AC Load-Equipment:
0
mounted hazardous
AC Load-Lighting:
1.6 W/SF
location fixture classified
Number of People:
0
for use. Chemical storage
--
room.
Notes:
--
Notes:
1. Provide seal off fitting
as required.
PLUMBING AND MEDICAL GASES
POWER
Coordinate with SFO; refer to Template SFO
Section 6.6 and Schedule B
Coordinate with SFO; refer to Template SFO
Section 6.7, 6.7.1, 6.7.3, 6.7.4 and Schedule B.
Cold Water:
--
Hot Water:
--
General:
Yes
Laboratory Air:
--
Emergency:
--
Laboratory Vacuum:
--
Notes:
1. Provide hazardous
Sanitary Drain:
--
classified wiring device
Reagent grade Water:
--
for use. Chemical
Medical Air:
--
storage room. Provide
Medical Vacuum:
--
seal off fittings as
Oxygen:
--
required.
Notes:
--
COMMUNICATION/SPECIAL SYSTEMS
Coordinate with SFO; refer to Template SFO
Sections 6.7.8, 6.7.9, and Schedule B.
Data:
--
Telephone:
--
Intercom:
--
Nurse Call:
--
Public Address:
--
Radio/Entertainment: --
MATV:
--
CCTV:
--
MID:
--
VA DESIGN GUIDE
APPENDIX B
LEASE BASED OUTPATIENT CLINICS
3-12.3