September 30, 1995
Design Standards
ARCHITECTURAL
Floor Area
42 m (457 ft)
Wall Finish
gwb
Ceiling
at
Base
rb
Ceiling Height
2700 mm (9 ft)*
Floor Finish
vct
RFI Shielding
yes
Slab Depression
-
Soundproofing
-
Floor Load
-
*4100 mm (13'6") min over magnet 1.5 m (16 ft) min
SPECIAL EQUIPMENT
Lighting
ELECTRICAL
Power
General
750 lx (70 fc)
General
-
Special
-
Emergency
ventilation system
Direct current, non-ferrous, incandescent fixtures with
Copper cladding on wiring
PAR38 halogen flood lamps on dimmer switches
Telephone
yes
ADP
yes
HEATING, VENTILATING AND AIR CONDITIONING
AC Load Lights
2760 W
Dry Bulb Temp Cooling
23C (74F)
AC Load Equipment
70 W/m (6.5 W/ft)**
22C (72F)
Number of People
3
Relative Humidity - Cooling
50%
Noise Criteria
NC-40
Relative Humidity - Heating
30%
Room Pressure
(0)
Minimum % Outside Air
100
Min Air Changes per Hour 6
yes
Steam
-
yes***
**Minimum requirement
***Use only copper piping and aluminum or PVC ductwork inside the room.
PLUMBING AND LABORATORY GASES
Sanitary Drain
-
Cold Water
yes
Acid Waste
yes
Hot Water
yes
Other
-
Reagent Grade Water
-
Fuel Gas
yes
Laboratory Air
yes
-
Laboratory Vacuum
yes
Additional Notes:
1. Utility requirements for magnets vary greatly among manufacturers. There are different kinds of magnets with
specific utility requirements. Requirements also vary if gradient power supply and gradient coils are used, and
whether they are air-cooled or water-cooled. A water-cooled system, consisting of a closed loop water supply with
water-to-water heat exchanger to a refrigerator type heat exchanger, is recommended for gradient power supply
and gradient coils. The water must be filtered to protect equipment. Provide domestic water back-up for cooling
water in case of emergency.
Guide Plate 5-2
VA Design Guide -- Research Laboratory