Final Draft 10-18-96
Revised February 1997
Equipment Guide List
Ambulatory Surgery Operating Room
SYMBOL
QTY AI
DESCRIPTION
1
CC
WINDOW, VIEWING, LEAD GLASS, FOR PATIENT OBSERVATION
(PG-18-1, MCS 13091; H-08-4, SD 24G)
20
CC
RECEPTACLE, ELECTRICAL, SIMPLEX, 120 VOLT, 20 AMP,
ONE GROUNDING JACK PER MODULE AND FOUR SIMPLEX
(PG-18-1, MCS 16140; H-08-3, CS 801-3)
4
CC
ILLUMINATOR, FILM, X-RAY, RECESSED, 120 VOLT, 20 AMP,
356 mm X 432 mm (14" X 17") (INSTALLATION NOT TO BE
COMBINED WITH IPU'S OR OTHER ELECTRICAL DEVICES)
(PG-18-1, MCS 16510)
AR
CC
CONTROL (PG-18-1, MCS 16510)
2
CC
COLUMN, TELESCOPING OR ARTICULATING, CEILING
MOUNTED (PG-18-1, MCS 15491)
COLUMN A : LOCATE AT HEAD OF TABLE,1200 mm TO 1800 mm (48"
TO 72") FROM THE CENTERLINE AND 600 mm TO 1200 mm
(24" TO 48") TO THE LEFT OF THE CENTERLINE OF THE TABLE
COLUMN B : LOCATE AT THE FOOT OF THE TABLE,1200 mm TO 1800
mm (48" TO 72") FROM THE CENTERLINE AND 600 mm TO 1200 mm
(24" TO 48") TO THE RIGHT OF THE CENTERLINE OF THE TABLE.
EACH COLUMN CONTAINS THE FOLLOWING:
2 INLETS, VACUUM (PG-18-1, MCS 15491)
1 OUTLET, NITROUS OXIDE (PG-18-1, MCS 15491)
2 OUTLET, OXYGEN (PG-18-1, MCS 15491)
1 OUTLET, MEDICAL AIR (PG-18-1, MCS 15491)
1 INLET, VACUUM, DEDICATED ANESTHESIA GAS
EVACUATION (PG-18-1, MCS 15491)
1 INLET, MASS SPECTROMETER (BLANK OUTLET)
2
CC
ISOLATED POWER UNIT PROVIDES ISOLATED ELECTRICAL
POWER, INCLUDES LINE ISOLATION MONITOR, ISOLATION
FROM ELECTRICAL SERVICE)
1
CF
LIGHT, MAJOR, SURGICAL WITH VARIABLE INTENSITY
CONTROL, SINGLE POINT SUSPENSION, CEILING MOUNTED
(LOCATION INSTRUCTIONS AVAILABLE FROM ELECTRICAL
SERVICE) (PG-18-1, MCS 16515)
AR
VV
TABLE, OPERATING, MOBILE
1
CC
CLOCK, ELECTRIC WITH SWEEP SECOND HAND, RECESSED
(PG-18-1, MCS 10360)
VA DESIGN GUIDE
AMBULATORY SURGERY
PAGE 4-1